topbanner

Our Products

Contact Us Online

Mechanical Pipettes

Mechanical Pipettes (Adjustable and Fixed Volume)

Single-channel Adjustable Volume Pipettes

Cat No.

Volume Range

Increment

Test Volume 

Maximum permissible systematic error (Inaccuracy)

Maximum permissible random error (Imprecision)

 

%

μl

%

μl

L061001

0.1-2.5μl

0.05μl

2.5

2.50%

0.0625

2.00%

0.05

1.25

3.00%

0.0375

3.00%

0.0375

0.25

12.00%

0.03

6.00%

0.015

L061002

0.5-10μl

0.1μl

10

1.00%

0.1

0.80%

0.08

5

1.50%

0.075

1.50%

0.075

1

2.50%

0.025

1.50%

0.015

L061003

2-20μl

0.5μl

20

0.90%

0.18

0.40%

0.08

10

1.20%

0.12

1.00%

0.1

2

3.00%

0.06

2.00%

0.04

L061004

5-50μl

0.5μl

50

0.60%

0.3

0.30%

0.15

25

0.90%

0.225

0.60%

0.15

5

2.00%

0.1

2.00%

0.1

L061005

10-100μl

1μl

100

0.80%

0.8

0.15%

0.15

50

1.00%

0.5

0.40%

0.2

10

3.00%

0.3

1.50%

0.15

L061006

20-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

20

3.00%

0.6

1.00%

0.2

L061007

50-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

50

1.00%

0.5

0.40%

0.2

L061008

100-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

100

2.00%

2

0.70%

0.7

L061009

200-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

200

0.90%

1.8

0.30%

0.6

L061010

1000-5000μl

50μl

5000

0.50%

25

0.15%

7.5

2500

0.60%

15

0.30%

7.5

1000

0.70%

7

0.30%

3

L061011

2-10ml

0.1ml

10ml

0.60%

60

0.20%

20

5ml

1.20%

60

0.30%

15

2ml

3.00%

60

0.60%

12

   

8-channel Adjustable Volume Pipettes

Cat No.

Volume Range

Increment

Test Volume 

Maximum permissible systematic error (Inaccuracy)

Maximum permissible random error (Imprecision)

 

%

μl

%

μl

L061201

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15

5

2.50%

0.125

2.50%

0.125

1

4.00%

0.04

4.00%

0.04

L061202

5-50μl

0.5μl

50

1.00%

0.5

0.50%

0.25

25

1.50%

0.375

1.00%

0.25

5

3.00%

0.15

2.00%

0.1

L061203

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75

150

1.00%

1.5

0.50%

0.75

50

1.50%

0.75

0.80%

0.4

12-channel Adjustable Volume Pipettes

Cat No.

Volume Range

Increment

Test Volume 

Maximum permissible systematic error (Inaccuracy)

Maximum permissible random error (Imprecision)

 

%

μl

%

μl

L061301

0.5-10μl

0.1μl

10

1.50%

0.15

1.50%

0.15

5

2.50%

0.125

2.50%

0.125

1

4.00%

0.04

4.00%

0.04

L061302

5-50μl

0.5μl

50

1.00%

0.5

0.50%

0.25

25

1.50%

0.375

1.00%

0.25

5

3.00%

0.15

2.00%

0.1

L061303

50-300μl

5μl

300

0.70%

2.1

0.25%

0.75

150

1.00%

1.5

0.50%

0.75

50

1.50%

0.75

0.80%

0.4

 

Fixed Volume Pipettes

Cat No.

Volume Range

Increment

Test Volume 

Maximum permissible systematic error (Inaccuracy)

Maximum permissible random error (Imprecision)

 

%

μl

%

μl

L062001

5μl

-

5μl

1.30%

0.065

1.20%

0.06

L062002

10μl

-

10μl

0.80%

0.08

0.80%

0.08

L062003

20μl

-

20μl

0.60%

0.12

0.50%

0.1

L062004

25μl

-

25μl

0.50%

0.125

0.30%

0.075

L062005

50μl

-

50μl

0.50%

0.25

0.30%

0.15

L062006

100μl

-

100μl

0.50%

0.5

0.30%

0.3

L062007

200μl

-

200μl

0.40%

0.8

0.20%

0.4

L062008

250μl

--

250μl

0.40%

1

0.20%

0.5

L062009

500μl

-

500μl

0.30%

1.5

0.20%

1

L062010

1000μl

-

1000μl

0.30%

3

0.20%

2

L062011

2000μl

-

2000μl

0.30%

6

0.15%

3

L062012

5000μl

-

5000μl

0.30%

15

0.15%

7.5